Quality Control · Radiography

Radiography Quality Control Tests: What Is Measured in DR and CR Systems?

An X-ray unit can degrade without ever failing: the tube voltage drifts, the AEC terminates differently, or unnoticed artifacts appear on the detector. The job of quality control is to catch these before they reach patient images. What is measured, how, by whom, and how it is judged — with DR vs CR differences, grounded in IAEA and Bushberg.

An X-ray unit can degrade in image quality without ever failing: the tube voltage can drift from the set value, the AEC can terminate the exposure differently, or unnoticed artifacts can appear on the detector. The purpose of quality control (QC) tests is to catch these changes before they reach patient images. So QC is not only a physicist's concern: these tests sit behind the image a radiologist and technologist see every day, and for a physicist in radiotherapy or nuclear medicine it is valuable to see how diagnostic radiology is kept in check. In this article we take the radiography-specific tests one by one — what is measured, how, with what equipment, by whom, and how the result is judged — distinguishing DR (digital flat panel) from CR (storage phosphor) where relevant.

What is it?

In its simplest definition, quality control is the process of verifying, by measurement, that a radiography system's performance today is the same as it was yesterday. Technically, this means being able to show that every link of the imaging chain — generator, tube, collimator, AEC and digital detector — works within accepted limits. QC is the measurable, technical leg of a broader quality management system (QMS); the IAEA emphasizes that the medical physicist plays a central role in this system, especially in equipment performance.1

Tests are of two kinds: acceptance testing before the device enters service (establishing baseline/reference values), and the constancy tests repeated regularly thereafter.1

Why is it done?

Because an X-ray unit can drift over time without failing outright: calibration values can change, detector response can shift, the AEC can terminate differently. These drifts are invisible to the eye. QC has three aims: patient safety (avoiding unnecessary dose), diagnostic quality (consistent, reliable images) and regulatory compliance (mandatory performance limits). For example, minimum HVL is legally required in the US under 21 CFR 1020.30.23

What happens when it drifts?

Every test has a "what if it's not done / drifts" counterpart:

What is measured, and how?

Most generator and tube tests can be done with a non-invasive multimeter placed at the detector plane and a calibrated dosimeter.

1 · Tube voltage (kVp) accuracy. Using a non-invasive kVp meter at the detector plane, several kVp settings are exposed and the measured peak voltage is compared with the value set on the console. Deviation affects both contrast and dose.

2 · Radiation output: reproducibility and linearity. The reproducibility of the output across repeated exposures at the same technique is assessed via the coefficient of variation (COV);2 linearity with mAs is checked by whether the measured output rises proportionally across different mAs values. 21 CFR 1020.30 sets limits for both.3

3 · Beam quality: half-value layer (HVL). Aluminum filters of increasing thickness are placed in front of the beam and the dose is recorded at each thickness; the filter thickness that halves the initial dose is defined as the HVL. The measured HVL must meet the minimum HVL requirement for that kVp (Bushberg Table 6-3; 21 CFR 1020.30).23

4 · Automatic exposure control (AEC). The aim is to verify that even when phantom thickness or tube voltage changes, the exposure reaching the detector (and the associated image indicator) stays constant within acceptable limits.1 Measurements are made at different thicknesses (e.g. PMMA / aluminum slabs) and different kVp settings.

5 · Light field – beam field congruence and collimation. The collimator's light field must coincide with the actual X-ray field; per Bushberg the misalignment, along the length or width of the field, must not exceed 2% of the SID.2 Alignment of the beam field with the image receptor is also checked.

6 · Digital detector — DR. Two points stand out in DR:

7 · Storage phosphor — CR. CR cassettes need extra tests:

8 · Spatial resolution and low contrast. Resolution is assessed with a line-pair phantom (or by measuring the MTF); the limiting resolution is the spatial frequency at which the modulation drops to about 5% of its initial value — i.e. where the MTF curve practically reaches zero.1 Low contrast is examined with contrast-detail phantoms.

9 · Reject (repeat) analysis. The rate and causes of rejected/repeated images are tracked regularly; a rising repeat rate can be an early sign of a systematic problem in the AEC, technique factors or positioning.

collimatorimage receptor (detector)beam field (irradiated)light field (collimator light)misalignment ≤ 2% of SIDLight field ↔ beam field congruence
The dashed yellow frame is the collimator's light field; the filled blue area is the actual irradiated beam field. The misalignment between them must not exceed 2% of the SID along the length or width of the field.2

What equipment is needed?

Who does it, who checks?

In practice the division of labor is roughly:

How are results evaluated?

Each measurement is compared with the baseline value from acceptance testing and judged against predefined levels:

In DR the same logic applies to EI/DI and the reject rate: a marked deviation from the target EI (DI) or a rising repeat rate triggers a protocol review. These levels and test frequencies are set by national regulations and protocols (IAEA, AAPM, EUR); the values here are examples — the facility protocol and competent authority take precedence.5

The series continues
This is the first article in the modality-by-modality QC series. Next up: Computed Tomography, Mammography, Fluoroscopy and Nuclear Medicine. For the general framework: Why Quality Control Matters. For how radiography works: Radiography.

References

  1. IAEA. Diagnostic Radiology Physics: A Handbook for Teachers and Students (STI/PUB/1564), 2014. Bölüm 19 (Quality Management) — kalite yönetim sistemi, kabul/sabitlik testleri ve medikal fizik uzmanının rolü; §5.6.1 (kolimatör ve ışık alanı); dijital radyografi bölümlerinde ışınlama indeksi (exposure index) tekrarlanabilirliği ve tutarlılığı; sınır çözünürlüğün MTF≈%5 noktasında tanımı (Bölüm 4–5). iaea.org
  2. Bushberg JT, Seibert JA, Leidholdt EM, Boone JM. The Essential Physics of Medical Imaging, 3rd ed. Lippincott Williams & Wilkins, 2011. Işık–ışın alanı çakışması SID'in %2'si içinde (s.187); tekrarlanabilirlik / varyasyon katsayısı, COV (s.85); minimum HVL gereksinimleri (Tablo 6-3, 21 CFR 1020.30).
  3. U.S. FDA. 21 CFR 1020.30 — Diagnostic X-ray Systems: ışın kalitesi (minimum HVL), çıkış tekrarlanabilirliği/doğrusallığı ve ışık/ışın alanı gereksinimleri. ecfr.gov
  4. AAPM Task Group 116. An Exposure Indicator for Digital Radiography; IEC 62494-1. Standartlaştırılmış ışınlama indeksi (EI), hedef ışınlama indeksi (EIT) ve sapma indeksi (deviation index, DI) tanımları.
  5. Test setleri, tolerans değerleri ve sıklıklar ilgili ulusal mevzuata ve AAPM / EUR (European Guidelines) gibi protokollere göre belirlenir; bu yazıdaki değerler örnektir, kurum protokolü ve yetkili otorite esastır.
Note: This content is for education; for clinical decisions or regulatory compliance, consult a qualified medical physicist and current regulations.

← All articles