MRI produces a correct image only if two things are perfect: a uniform magnetic field across the whole field of view, and a clean radiofrequency (RF) signal. In reality neither is perfect. A metal implant bends the field, a tissue–air boundary erases signal, patient motion creates ghosts, and signal processing itself produces rings. This is the MRI deep-dive companion to our general artifacts guide.
Classifying artifacts
Bushberg divides MRI artifacts into three broad groups:1
- Machine-related: magnetic field inhomogeneity (global/focal), susceptibility, gradient and RF artifacts.
- Patient-related: motion (voluntary/involuntary) and flow (blood, CSF).
- Signal-processing-related: chemical shift, wraparound, ringing (Gibbs).
A non-uniform field causes tissues to be mismapped in the image; this is why shimming, self-shielded magnets and regular maintenance are the foundation of MRI quality assurance.1
Magnetic susceptibility
Magnetic susceptibility is the ratio of a tissue's induced internal magnetization to the external field. If susceptibility varies little across the FOV the field stays uniform; but a sharp change distorts it.1 The most common change occurs at tissue–air interfaces (lungs, sinuses), causing signal loss from faster T2* dephasing; any metal (ferrous or not) also creates marked distortion in nearby tissue.1 Interestingly, susceptibility is sometimes diagnostically helpful: the susceptibility effects of hemorrhage products (deoxyhemoglobin, methemoglobin, hemosiderin) help date a bleed as acute/subacute/chronic.1
Chemical shift
Protons in fat and water precess at slightly different (Larmor) frequencies. Because MRI reads position from frequency, this small frequency difference shifts fat by a few pixels relative to water along the frequency-encoding direction.1 The result is a dark band on one side of the fat–water boundary and a bright band on the other — classically seen at kidney or vertebral margins. The shift magnitude varies with field strength and the chosen bandwidth.
Wraparound
Anatomy outside the field of view (FOV) but within the slice volume is not ignored; it is mapped onto the opposite side.1 The cause is undersampling (aliasing): signals outside the FOV lie above the Nyquist limit and are interpreted as if they were low-frequency, "wrapping" to the far edge of the image. It is typically seen along the phase-encoding direction; reduced by enlarging the FOV or applying an anti-aliasing saturation pulse outside it.1
Motion and ghosting
Motion is MRI's most ubiquitous and noticeable artifact: voluntary/involuntary movement and flow (blood, CSF).1 The long acquisition time of some sequences raises the chance of motion blur and contrast loss. Motion artifacts appear mostly along the phase-encoding direction, because adjacent phase-encoding measurements in k-space are separated by a TR interval (which can be 3,000 ms or longer), so even slight motion changes the recorded phase over that time.1 The visual result is faint copies arrayed along the phase-encode direction — ghosting. The frequency-encoding direction is less affected. The simplest remedy is to swap the phase and frequency encoding gradients (PEG/FEG) to move the ghosts away from the region of interest.1
Ringing (Gibbs)
Ringing (the Gibbs phenomenon) is a set of faint parallel lines near sharp boundaries.1 Its cause is the lack of high-frequency signals in k-space needed to represent a sharp transition; so it is more pronounced at small matrix sizes (e.g. 128 instead of 256).1 It is typically seen at the skull–brain boundary, and increasing the matrix size is the most direct fix.
RF and machine
RF interference leaves its own patterns: narrow-band noise produces a zipper pattern perpendicular to the frequency-encoding direction, while broadband noise produces a herringbone pattern.1 The remedy is proper site planning and RF-shielding the room with a Faraday cage.1 Because these are machine/environment artifacts, their solutions lie on the engineering and quality-assurance side.
References
- Bushberg JT, Seibert JA, Leidholdt EM, Boone JM. The Essential Physics of Medical Imaging, 3rd ed. Lippincott Williams & Wilkins, 2011. §13.5 MR Artifacts (s.475–486): sınıflama — makine/hasta/sinyal işleme (s.475); manyetik duyarlılık (susceptibility) — doku–hava arayüzü ve metal, T2* faz kaybı (s.475–476, Şekil 13-24); hareket artefaktları — faz kodlama yönü, akış (kan/BOS), hayalet (ghosting), PEG/FEG değişimi (s.480, Şekil 13-29); kimyasal kayma (s.480); halkalanma/Gibbs — matris boyutu (s.484, Şekil 13-37); katlanma/wraparound (s.485–486, Şekil 13-38); RF zipper/herringbone ve Faraday kafesi (s.477–478, Şekil 13-44). Sayfa numaraları bu baskıya aittir.